Method
Cell Counting with a Hemocytometer
A hemocytometer turns a drop of suspension into a cells-per-millilitre number. The grid, the boundary rule and one multiplication are all there is to it.
Quick answer
To count cells on a hemocytometer, load about 10 µL of a trypan blue mixture under the coverslip, count the four large corner squares, then apply: cells per mL = mean count per square × dilution factor × 10,000. The factor of 10,000 comes from the chamber: each large square holds exactly 1 × 10⁻⁴ mL.
- Each large square holds 1 × 10⁻⁴ mL, hence the ×10,000.
- Count four corner squares and average them.
- Use one boundary rule: two edges in, two edges out.
- Target 100 cells or more in total, and dilute if squares are crowded.
The counting grid
An improved Neubauer chamber carries a 3 mm by 3 mm grid of nine large squares, each one square millimetre. The coverslip rests 0.1 mm above the grid, so every large square defines a fixed volume of 0.0001 mL. That fixed volume is what makes the count quantitative.
- Count cells touching these two edges
- Skip cells touching these two edges
The method, step by step
- Make a single-cell suspension. Trypsinise, neutralise and pipette gently until clumps are gone. Clumps are the largest source of counting error.
- Dilute with trypan blue. Mix equal volumes of suspension and 0.4 percent trypan blue, giving a dilution factor of 2, and count within a few minutes.
- Load the chamber. Place the coverslip, then touch about 10 µL to the edge so capillary action fills the chamber in one go. Do not overfill or refill.
- Count four corner squares. Count the cells inside each 1 mm square, applying one boundary rule: include cells touching the top and left edges, exclude the bottom and right.
- Average and apply the formula. Cells per mL equals the mean count per large square multiplied by the dilution factor and by 10,000.
- Calculate viability. Viability is live cells divided by total cells, times 100. Blue cells are dead, because they cannot exclude the dye.
Working the formula
Take a real example. You count four corner squares and get 82, 90, 78 and 90 cells.
mean per square = (82 + 90 + 78 + 90) / 4 = 85
cells per mL = 85 × 2 (trypan blue) × 10,000 = 1.7 × 10⁶
cells in 10 mL = 1.7 × 10⁷From there, seeding is simple division. To put 5,000 cells in each well of a 96-well plate in 100 µL, you need 5 × 10⁴ cells per mL, so dilute the suspension 34-fold. Volumes and areas for every format are on the surface area and volume chart, and target densities are in cell confluency.
Viability with trypan blue
Count live and dead cells separately as you go. Viability is live cells divided by total cells, as a percentage. A healthy, well-handled culture is usually above 90 percent, and a sudden drop points at over-trypsinisation, harsh pipetting or a culture left too long.
Trypan blue reports membrane integrity only. Cells that are dying but still intact count as live, so a high viability figure does not guarantee a healthy culture. For function, use a metabolic or ATP-based assay instead.
Where counts go wrong
- Clumps. A clump counts as one cell and hides several, which biases the result low. Break clumps before loading, never on the grid.
- Overfilling the chamber. Liquid in the moat lifts the coverslip and changes the depth, which changes the volume the whole calculation assumes.
- Counting too few cells. Below about 100 cells total, random variation alone moves the result by double digits.
- Letting the sample settle. Mix immediately before loading. Cells sediment in a tube within a minute.
- Inconsistent edge rule. Switching which edges you count between squares quietly inflates the total.
Frequently asked questions
What is the formula for cell counting with a hemocytometer?
Cells per mL equals the average number of cells in one large square, multiplied by the dilution factor, multiplied by 10,000. The 10,000 comes from the chamber geometry: each large square covers 1 mm by 1 mm and the coverslip sits 0.1 mm above it, so one square holds 0.0001 mL.
Why multiply by 10,000?
Because one large square holds a known, very small volume. One square millimetre of area and a depth of 0.1 mm give 0.1 cubic millimetres, which is 0.0001 mL. To scale a count from that volume to one millilitre you multiply by 10,000.
Which squares do you count on a hemocytometer?
For a typical cell suspension, count the four large corner squares of the Neubauer grid and average them. The centre square, subdivided more finely, is intended for very dense samples such as red blood cells. Counting four squares rather than one averages out uneven distribution.
How do you handle cells sitting on the grid lines?
Pick two adjoining edges and count only cells touching those, usually the top and the left, and skip cells touching the other two. Applied consistently, each cell is counted exactly once across the whole grid. Changing the rule midway through is a common source of inflated counts.
How many cells should I count for an accurate result?
Aim for roughly 20 to 50 cells per large square, and at least 100 cells in total across the squares you count. Fewer than that and the random error becomes large. If the squares are too crowded to count, dilute the sample further and recount rather than estimating.
What does trypan blue tell you?
Trypan blue stains cells whose membranes are damaged. Live cells with intact membranes exclude the dye and look bright, while dead cells take it up and appear blue. The stain measures membrane integrity, not metabolic health, and it becomes toxic over time, so count within a few minutes of mixing.
Why does my hemocytometer count differ from an automated counter?
Automated counters apply fixed size and shape gates, so they may exclude debris and small cells that a person would count, or include doublets as single events. Clumping affects both. Treat the two methods as separate assays and stay with one of them within an experiment.