Cell culture

Cell Culture Inserts

An insert turns one well into two compartments separated by a porous membrane. That single change supports barrier models, migration assays and co-culture.

Quick answer

A cell culture insert, often called a Transwell, is a cup with a porous membrane base that sits inside a well and creates an apical compartment above and a basolateral compartment below. Pore size sets the application: 0.4 µm for barrier and transport models, 3 µm for small migrating cells, and 5 to 8 µm for migration and invasion assays.

Key takeaways
  • An insert splits one well into apical and basolateral compartments.
  • Pore size decides whether cells can cross the membrane.
  • Keep medium levels balanced unless flow is intended.
  • Record the compartment, not just the well ID, on your plate map.

How inserts work

Cells are seeded on the membrane inside the insert and fed from both sides. Because the membrane is permeable, molecules move between compartments while the cells stay where they were placed. What you measure depends on which side you sample.

  • Barrier models. Intestinal, lung or blood-brain-barrier cells form a tight monolayer. You then measure how readily a compound crosses it.
  • Migration and invasion. Cells are placed above, a chemoattractant below, and the cells that cross the membrane in a set time are counted. Coating the membrane with a matrix gel turns migration into an invasion assay.
  • Co-culture. Two cell types share medium but never touch, which isolates soluble signalling from direct contact.

Choosing a pore size

Pore sizeUseWhy
0.4 µmBarrier and transport modelsCells cannot pass through, so the monolayer stays intact while molecules diffuse
1.0 µmCo-culture with soluble signalling onlyBlocks most cell movement while allowing good medium exchange
3.0 µmMigration of small cells such as lymphocytesLarge enough for active migration, small enough to prevent passive fall-through
5.0 to 8.0 µmMigration and invasion of tumour and epithelial cellsStandard for chemotaxis assays; 8 µm is the usual default

Membrane material matters too. Polycarbonate is the general-purpose choice and is translucent. Polyester is clearer and better for imaging in place. PTFE is used where chemical resistance matters. Pore density changes the effective area for diffusion, so compare like with like when switching supplier.

Volumes by plate format

Inserts are made to fit standard plates, so the plate format sets the insert diameter and the volumes on each side:

PlateInsert diameterApicalBasolateralTypical use
6-well24 mm1.5–2.5 mL2.0–3.0 mLBarrier models needing plenty of material for harvest
12-well12 mm0.5 mL1.5 mLRoutine barrier and transport work
24-well6.5 mm0.1–0.3 mL0.6–0.9 mLMigration and invasion assays, the most common choice
96-well4.3 mm50–75 µL150–235 µLScreening, usually with automated handling

Nominal values; exact volumes differ between suppliers and insert heights. Plate dimensions behind these figures are on the surface area and volume chart.

Keep the levels even

A difference in liquid height between the two compartments creates hydrostatic pressure and drives flow through the membrane, which can push cells off it or distort a transport measurement. Fill the lower compartment first, then the insert.

Mapping a two-compartment plate

An insert experiment breaks the usual assumption that one well equals one sample. Well B3 now holds two samples that can be measured separately, so a results file labelled only by well ID is ambiguous.

  • Add a compartment column to your layout, with values such as apical and basolateral.
  • Keep the wording identical in the plate map and the reader export, so the two join cleanly.
  • Note the insert lot and pore size next to the plate ID, since membrane batch affects barrier formation.
  • Record TEER readings against the same well IDs, so barrier quality and endpoint data stay linked.

Build the base layout in the plate designer, export the CSV, then add the compartment column. The 24-well format, the most common for inserts, has its own template and worked layout.

Frequently asked questions

What is a cell culture insert?

A cell culture insert is a small cup with a porous membrane base that hangs inside the well of a standard plate. It splits one well into two compartments: the apical space inside the insert and the basolateral space in the well below. Cells grow on the membrane, with medium on both sides, which lets you study transport, migration and signalling between compartments.

What is a Transwell?

Transwell is Corning's brand name for its permeable supports, and it is used colloquially for inserts of any make. Other suppliers sell the same concept under their own names. They are interchangeable in principle, but membrane material, pore density and insert height differ, so values are not directly comparable between brands.

Which pore size should I choose?

Match it to whether cells should pass through. Use 0.4 µm for barrier and transport models, where the monolayer must stay intact. Use 3 µm for small migrating cells such as lymphocytes, and 5 to 8 µm for tumour and epithelial cell migration or invasion. Pores that are too large let cells drop through without migrating.

How much medium goes in an insert?

It depends on the format. A 6.5 mm insert in a 24-well plate takes 100 to 300 µL on top and 600 to 900 µL below. A 12 mm insert in a 12-well plate takes about 0.5 mL above and 1.5 mL below. Keep the two levels balanced unless the protocol calls for hydrostatic pressure, because a height difference drives flow through the membrane.

What is TEER and why measure it?

TEER is transepithelial or transendothelial electrical resistance, measured in ohms times square centimetre. It is the standard non-destructive check that a barrier has formed and stayed intact: resistance rises as tight junctions close. Measure it at the same temperature every time, because readings fall as the plate cools.

Can I image cells on an insert membrane?

Yes, but plan for it. Standard membranes are translucent and scatter light, so most people cut the membrane out and mount it on a slide for high-resolution imaging. Optically clear polyester membranes allow imaging in place at lower resolution, which is usually enough for confluency checks.

How do I record an insert experiment on a plate map?

Treat each well as two entries, because it holds two distinct volumes. Note on the map which compartment a sample came from, apical or basolateral, and keep the same wording in the exported file. A reading labelled only by well ID is ambiguous once inserts are involved.