Measurement
OD600
Optical density at 600 nm is how most labs track bacterial growth. It is quick, it needs no reagents, and it is not quite the absorbance measurement it appears to be.
Quick answer
OD600 is the optical density of a culture at 600 nm, used as a stand-in for cell density. Bacteria scatter light rather than absorb it, so the value depends on the instrument and on the light path. In a microplate the path is set by the fill volume, roughly 0.31 cm for 100 µL, so plate readings run about a third of cuvette readings for the same culture.
- OD600 measures scattering, so it is instrument dependent.
- Plate values are lower than cuvette values for the same culture.
- Stay below roughly 0.6 for a near-linear response.
- Blank with sterile medium at the same volume, on the same plate.
What OD600 actually measures
A plate reader or spectrophotometer sends light through the sample and records how much reaches the detector. With a dye, the missing light has been absorbed. With a bacterial culture, almost none of it is: cells deflect photons out of the beam, and the instrument counts them as lost. The number is still useful, because more cells scatter more light, but it is a relative measure rather than a physical constant.
Two consequences follow. Different instruments give different values for the same tube, because their optics collect scattered light differently. And an OD600 to cells per millilitre conversion only holds for the organism, growth phase and instrument it was measured on.
OD600 in a microplate
Reading in a plate is faster, uses less culture, and lets a reader follow dozens of wells in parallel with shaking and temperature control. The trade is the light path. A cuvette holds it at 1 cm, while a well gives you whatever depth you pipetted.
| Vessel and volume | Approximate light path | Reading vs cuvette |
|---|---|---|
| Cuvette | 1.0 cm | 1.00× |
| 96-well, 100 µL | ≈ 0.31 cm | ≈ 0.31× |
| 96-well, 200 µL | ≈ 0.62 cm | ≈ 0.62× |
| 48-well, 500 µL | ≈ 0.45 cm | ≈ 0.45× |
| 24-well, 1 mL | ≈ 0.53 cm | ≈ 0.53× |
Path length is the fill volume divided by the well’s base area, using the figures on the surface area chart. Meniscus shape makes the real value slightly different, which is another reason to keep volumes identical.
Two wells with the same culture but different fill volumes give different OD600 values. Pipette the same volume everywhere, including blanks, or the growth curves will not be comparable.
Running a growth curve in a plate
- Use a 48 or 96-well plate with a lid and fill the outer wells with medium if you are not using them, to slow evaporation over a long run.
- Inoculate to a low starting OD, around 0.01 to 0.05, so the exponential phase is captured rather than missed.
- Shake between reads to keep cells suspended and oxygen available. Without shaking, cells settle and the curve flattens for the wrong reason.
- Read every 10 to 30 minutes at 600 nm, with the correction off, and hold the temperature constant.
- Reserve blank wells of sterile medium, and treat any rise there as contamination.
- Fit the growth rate on log-transformed values within the exponential phase only.
A worked plate layout for six isolates across seven conditions, with sterile blanks in the last column, is on the 48-well plate template.
Reading the curve
| Phase | Typical OD600 | What is happening | Practical note |
|---|---|---|---|
| Lag | below ~0.05 | Cells adapt to the medium and do not divide yet | Length depends on the inoculum's previous state, not only its size |
| Exponential | ~0.05 to 0.6 | Division at a constant rate, and the only phase where growth rate is meaningful | Fit growth rate here, on log-transformed values |
| Late exponential | ~0.6 to 1.0 | Readings start to compress as light scatters more than once | Dilute before reading, or expect an underestimate |
| Stationary | plateau | Division balances death as nutrients run out | Final OD depends on medium, not only on the organism |
Frequently asked questions
What does OD600 mean?
OD600 is the optical density of a culture measured at a wavelength of 600 nm. It is used as a proxy for how many cells are present: the denser the culture, the less light reaches the detector. Six hundred nanometres is chosen because cell pigments and most media absorb little there, so the signal is dominated by the cells themselves.
Is OD600 an absorbance measurement?
Not really, even though instruments report it as absorbance. Bacteria scatter light rather than absorb it, so the photons are deflected out of the beam instead of being taken up. That is why OD600 readings depend on the instrument's optical geometry, and why two machines can give different values for the same tube.
How many cells are in an OD600 of 1?
For Escherichia coli in a standard cuvette, an OD600 of 1 corresponds very roughly to 8 × 10⁸ cells per mL, but published values range several fold. It depends on the strain, the growth phase, cell size and the instrument. Treat any conversion factor as something you calibrate yourself by plating for colony counts.
Why is OD600 lower in a microplate than in a cuvette?
A cuvette fixes the light path at 1 cm, while in a plate the light passes only through the depth of liquid in the well, roughly 0.31 cm for 100 µL. The same culture therefore reads about a third as high in a plate. Multiply by the path length, or use the reader's correction, if you need values comparable to cuvette readings.
Should I use path length correction for growth curves?
Usually not. The correction measures water in the near infrared, and dense cultures scatter that light too, which inflates the estimate. For growth curves it is normal to leave the correction off, keep the fill volume identical in every well, and compare values only within the run.
What is the usable range of OD600?
Readings stay close to linear up to roughly 0.6 in a cuvette. Above that, light scattered by one cell is scattered again by another, so the measured value rises more slowly than the true cell count. Dilute dense cultures into the linear range and multiply back, rather than reading a thick sample directly.
How do I blank an OD600 reading in a plate?
Use wells of sterile medium on the same plate, with the same volume as the samples, and subtract their mean. A separate blank plate will not capture that plate's lid, plastic and fill level. Include several blanks so contamination in one is obvious.