Template library · 6 plate formats · 6 file types

Well Plate Templates

Blank, standards-accurate microplate templates for every common plate format. This page also shows worked layouts for qPCR, MIC testing, siRNA knockdown and 384-well screening, and which template fits deep-well, PCR and strip plates.

Quick answer

A well plate template is a blank, labelled copy of a microplate grid for mapping samples, controls and standards to wells before an experiment. Free templates are available here for the six standard SBS/ANSI formats: 6-well (A1–B3), 12-well (A1–C4), 24-well (A1–D6), 48-well (A1–F8), 96-well (A1–H12) and 384-well (A1–P24). Each comes as PDF, Word, Excel, PowerPoint, PNG and SVG, free for research, teaching and commercial lab use.

Blank templates by plate format

Every template shares the 127.76 × 85.48 mm plate footprint, so the only difference is the grid. Each card shows a plan that fits that grid evenly. Download a file directly, or open the format page for dimensions, volumes and seeding guidance.

Starter layouts for common assays

A blank template becomes useful once controls and replicates are placed deliberately. The layouts below are widely used starting points drawn on the templates above. They are not validated protocols, so adapt them to your assay kit, the relevant CLSI or EUCAST document, or your lab’s SOP. You can rebuild any of them in the plate designer and export a PNG or CSV.

qPCR plate layout on the 96-well template

96-well qPCR plate layout: target gene in rows A–D and reference gene in rows E–H, each with 14 samples in triplicate plus no-RT and no-template controls123456789101112ABCDEFGH
  • Target gene · cDNA samples 1–14
  • Reference gene · same 14 samples
  • No-RT control
  • No-template control (NTC)
Fig. 1. Sample-maximisation qPCR layout: 2 primer pairs × 16 triplicate slots = 96 wells.

Each primer pair gets half the plate. Rows A–D hold the target gene and rows E–H the reference (housekeeping) gene, such as GAPDH or ACTB. Every cDNA sample runs as a technical triplicate in three adjacent wells, which leaves room for 14 samples per gene. Running every sample for both genes on the same plate is called a sample-maximisation design, and it keeps run-to-run variation out of the ΔΔCq comparison.

  • No-template control (NTC): water instead of cDNA for each primer pair. Amplification here points to reagent contamination or primer-dimers.
  • No-RT control: RNA taken through reverse transcription without the enzyme. A signal here means genomic DNA is being amplified.
  • Relative quantification: the 2−ΔΔCq method assumes both assays amplify with similar, near-100% efficiency, so check efficiency with a standard curve first.

The same pattern scales to the 384-well template, where 5–20 µL reactions allow four primer pairs across the same samples. Record Cq thresholds and reaction volumes alongside the map, as the MIQE guidelines recommend.

Broth microdilution (MIC) plate layout

96-well MIC broth microdilution layout: two-fold antimicrobial dilutions in columns 1–10, growth control in column 11, sterility control in column 12, QC strain in row H123456789101112ABCDEFGH
  • Test isolates · two-fold dilutions, 64 → 0.125 µg/mL
  • QC reference strain
  • Growth control (no antimicrobial)
  • Sterility control (broth only)
Fig. 2. One isolate per row; colour intensity falls with antimicrobial concentration.

The minimum inhibitory concentration (MIC) is the lowest antimicrobial concentration that prevents visible bacterial growth. In the reference broth microdilution method described in CLSI M07 and ISO 20776-1, columns 1–10 carry a two-fold dilution series in cation-adjusted Mueller–Hinton broth, for example 64 down to 0.125 µg/mL. Rows A–G each take one test isolate at a final inoculum of about 5 × 105 CFU/mL.

  • Column 11, growth control: inoculum without antimicrobial. The run is invalid if these wells do not grow.
  • Column 12, sterility control: broth only. Any turbidity means the medium was contaminated.
  • Row H, QC strain: a reference organism such as Escherichia coli ATCC 25922 or Staphylococcus aureus ATCC 29213, whose MIC must fall in its published range.

Plates are typically read after 16–20 hours at 35 °C. Recording the concentration for each column on the template makes the MIC a direct read-off rather than a calculation.

384-well screening plate with control columns

384-well screening plate layout: negative controls in columns 1 and 24, positive controls in columns 2 and 23, test compounds in columns 3–22123456789101112131415161718192021222324ABCDEFGHIJKLMNOP
  • Negative control · vehicle (e.g. DMSO)
  • Positive control · reference inhibitor
  • Test compounds · 320 wells
Fig. 3. 64 control wells (4 columns × 16 rows) bracket 320 single-point test wells.

Primary screens usually test each compound once, so plate quality depends on the controls. Placing a negative and a positive control column on both sides of the plate exposes left-to-right drift from dispensing or reading. With 32 wells of each control, you can calculate the Z′-factor for every plate:

Z′ = 1 − 3(σpos + σneg) / |μpos − μneg|

A Z′ of 0.5 or higher indicates an excellent separation between controls. Plates below your cut-off are repeated before hits are called, and confirmed hits then move to dose-response testing for an IC50. Outer rows A and P are the most exposed to evaporation, as covered in common well plate template mistakes.

siRNA knockdown on the 24-well template

24-well siRNA knockdown layout: one condition per column (untreated, mock, non-targeting siRNA, two target siRNAs, positive-control siRNA) with four replicate wells each123456ABCD
  • Untreated cells
  • Mock · transfection reagent only
  • Non-targeting siRNA
  • Target siRNA · two independent sequences
  • Positive-control siRNA
Fig. 4. Six conditions × four wells; rows A–C for RNA, row D for a viability check.

Each column is one condition, which keeps pipetting simple and makes the map easy to read at the bench. The mock column separates effects of the transfection reagent from effects of the siRNA. The non-targeting siRNA is the baseline for calculating knockdown. Using two independent siRNA sequences against the same gene helps rule out off-target effects. A positive-control siRNA confirms that transfection worked.

Rows A–C give three wells for RNA extraction and RT-qPCR, typically 24–48 hours after transfection. Row D is kept for a viability or imaging check. Protein knockdown by western blot usually needs 48–72 hours.

Plans for 6-, 12- and 48-well templates

  • 6-well, protein or RNA harvest: control cells in A1–A3 and treated cells in B1–B3 give biological triplicates with enough lysate per well for a western blot.
  • 12-well, scratch (wound-healing) assay: four conditions in columns 1–4 with three wells each. Mark reference lines on the plate underside so the same fields are imaged at 0 hours and at each later timepoint.
  • 48-well, crystal-violet biofilm screen: seven strains in columns 1–7 and broth-only controls in column 8, with six replicate wells per column.

Which template fits your labware?

Many plates look different but share a standard grid. Choose the template by the grid, not by the plastic, well depth or skirt. The table covers the common cases, and the deep-well vs standard 96-well guide explains how those two plates differ in use.

LabwareGridTemplate to useNote
Deep-well plate, 1–2.2 mL polypropylene8 × 1296-well templateSame A1–H12 addresses; record the well volume.
PCR plate, skirted or semi-skirted8 × 1296-well templateNon-skirted plates keep the same grid.
ELISA strip plate in a frame8 × 1296-well templateLog the strip lot for each column or row.
Half-area or low-volume 96-well plate8 × 1296-well templateSame pitch, smaller wells.
96-position tube rack (e.g. 2D-barcoded tubes)8 × 1296-well templateMap tube barcodes to rack positions.
384-well PCR or assay plate16 × 24384-well template4.5 mm pitch; A1–P24.
24-well deep-well block, ~10 mL4 × 624-well templateCommon for bacterial and yeast cultures.
48-well deep-well block, ~5 mL6 × 848-well templateExpression cultures and small-scale growth.
Plate with cell-culture insertsParent plateSee noteUse the 6-, 12- or 24-well template; mark apical and basolateral contents.
1536-well plate32 × 48See noteNo printable template; manage as a well-ID list.

A 1536-well plate is the one standard format without a printable template. Its 32 rows run from A to Z and then AA to AF, across 48 columns at 2.25 mm pitch. At that density, layouts live in software as lists of well IDs rather than on paper. Dishes, flasks and chamber slides do not follow the microplate grid, so a well plate template does not apply to them.

Frequently asked questions

Which well plate template matches the plate on my bench?

Count the rows and columns, or check the catalogue listing. A 2 × 3 grid is a 6-well plate, 3 × 4 is 12-well, 4 × 6 is 24-well, 6 × 8 is 48-well, 8 × 12 is 96-well and 16 × 24 is 384-well. The bottom-right well gives the same answer: B3, C4, D6, F8, H12 or P24. Deep-well, PCR and strip plates take the template for their well count.

Is there a qPCR plate template?

Use the 96-well template for standard 0.1 or 0.2 mL qPCR plates and the 384-well template for 384-well qPCR plates, because the grids are identical. A qPCR layout is defined by what you add to the grid: technical triplicates, a no-template control (NTC) for each primer pair, a no-RT control and a reference gene on the same plate.

Can I use the 96-well template for a PCR plate, deep-well plate or tube rack?

Yes. Skirted and semi-skirted PCR plates, 1–2 mL deep-well plates, ELISA strip plates and 96-position tube racks all share the 8 × 12 grid, 9 mm well pitch and A1–H12 addresses. Only the well volume differs, so write the plate type and volume on the sheet.

Is there a 1536-well plate template?

Not as a printable file. A 1536-well plate has 32 rows (A–Z, then AA–AF) and 48 columns at 2.25 mm pitch, which is too dense to fill in by hand. Layouts at that density are kept as well-ID lists in a spreadsheet, LIMS or liquid-handler software. Each 384-well position corresponds to a 2 × 2 block of 1536-well positions.

Can I import a template layout into plate-reader or liquid-handler software?

Instrument software usually expects a list of well IDs with a sample name, concentration or volume against each, not a picture of a plate. Build the layout in the plate designer and export CSV to get that list, or type into the Excel template and convert the grid to one row per well. Match the instrument’s well-ID style, such as A1 or A01, so the files join cleanly.

Can I edit the templates and use them in SOPs, courses or publications?

Yes. You may download, edit, print and use the templates in research, teaching and commercial laboratory work, including SOPs, lab manuals, course handouts, electronic lab notebooks and figure panels. No sign-up or attribution is needed. The one restriction in our terms is that the templates themselves may not be resold or presented as your own product.